Description
After importing test data created in NOAH 2, the following error message appears:
No test(s) was found in the Speech Palette with the following measured conditions: Threshold, Air, Conductor, Numerals, Unaided
Solution
- Record the test conditions from the error message. In the above example they are Speech test, Threshold, Air Conductor, Numerals, Unaided. Then click OK until the message disappears.
- In NOAH System, click on Tools > Global Options.
- Select the Speech Tests tab and click New.
- Create a new test based on the parameters recorded in Step 1: Enter a name for this test at Notation. Next select the parameters, for example, Test type = SRT, Transducer Type = Phones, Aided Condition = Unaided, Word List = Numeric. Then click OK.
- The new speech test will now appear in the Speech Tests tab list. Click OK and exit.